II. TBI-DOSE PRESCRIPTION

Target volume.

Organs at risk.

Specification point for absorbed dose.

Prescribed dose to organs at risk.

Dose homogeneity.

Time dose distribution.


RECOMMENDATIONS - MEDICAL TREATMENT PLANNING

  1. As a general rule, the physician should prescribe the dose to the target volume and lung, and the timing schedule of the treatment.
  2. The target volume is the whole body including the skin, if not prescribed differently.
  3. A homogenous distribution of dose in the target volume is required.
  4. The reference dose to the target volume must be specified as the total dose delivered at the mid point of the abdomen at the level of the umbilicus.
  5. The dose to organs at risk must be specified. It is recommended neither to exceed a certain threshold dose nor to be far below this value.
  6. The lung dose should be specified at a point which is representative for more than 50% of the lung volume.
  7. The mean lung dose rate must be specified. Shielding of the lungs must be used in every session in order to have the same mean lung dose rate in each session.
  8. The estimated relative volume of lung receiving a reduced dose (by partial shielding) must be stated.
  9. The maximum and minimum values of dose in the transverse section containing the specification point should be within ±10%. The same criterion for dose homogeneity applies to the maximum and minimum values along the patient's midline.

image

Due to the large fields used, TBI techniques require a different approach than in standard Radiotherapy.

image

The particular TBI technique is imposed by the facilities of the Institution. A beam spoiler is recommended when high energy is used, to achieve dose homogeneity at shallow depths.


[ Back ] [ Index ] [ Next ]